Problem
Manual leak and electrical checks for electrosurgical probes created operator variability and risk of inconsistent defect detection.
02
Constraints
- Balloon leak detection required controlled inflation and decay monitoring.
- Electrical impedance and thermocouple response needed automated pass/fail criteria.
- Fixture and valve states had to prevent false failures and equipment damage.
- The station needed scalable production throughput.
03
Engineering Decisions
- Implemented an automated vapor generator sequence for leak and electrical checks.
- Used programmed pass/fail logic to remove subjective interpretation.
- Added valve-position checks and error-clearing logic to protect the test sequence.
- Structured output data around traceable serialized-unit results.
04
Validation
- Executed automated balloon inflation and decay checks.
- Verified impedance and thermocouple response against limits.
- Confirmed station interlocks and error states before production use.
05
Risk & Mitigation
- Micro-leaks could escape if decay limits were poorly defined; mitigated with automated pressure monitoring.
- False failures could occur from valve state errors; mitigated through state checks and error handling.
- Electrical faults could be missed by manual measurement; mitigated with automated sequence control.
06
Outcome
Automated inspection reduced operator variability, improved defect detection consistency, and supported traceable finished-device acceptance.